IL2560 – REVEXmax W

Multifunctional digital measuring device designed for tests of electrical appliances with movable and fixed power supply and also for inspections of arc welding equipment according to the ČSN EN 60974-4 standard requirements. The device can also measure the output voltage of SELV/PELV power supplies.
The device allows you to test electrical appliances both in terms of the electrical safety and their operational characteristics.
The device has a memory for storing measurement results with the possibility of transferring them to PC.

The price includes the calibration certificate.

34 965  (28 897  excl. VAT)

The price is valid for the online shop only

Measurement

  • Resistance of the protective conductor with a current of > 1 A
  • Insulation resistance by voltage of 500 V and 250 V
  • Leakage currents: by the substitude method, by the protective conductor and by the differential method
  • Leakage currents using a three-phase adapter or clamp meter
  • Apparent power / current drawn by the appliance
  • Apparent power / current using a clamp transformer
  • Mains voltage
  • Open circuit output voltage for arc welding systems
  • Voltage by means of measuring tips

Other features

  • Alternating quantities are measured using the TRMS method
  • Possibility to connect additional devices for measurement: three-phase adapters, clamp meters, etc.
  • Graphic OLED display (easy to read in different lighting conditions)
  • Up to 16,000 appliance can be stored in the internal memory
  • USB connector for connecting a barcode reader or RFID chip reader
  • Data transfer to PC via USB cable
  • "Help" function available directly in the device

The device is equipped with a number of control and safety functions

  • Checking the PE pin connection in the power socket before starting the measurement
  • Checking for dangerous touch voltage on the PE pin in the power socket before starting the measurement
  • Checking the amount of leakage current during measurement
  • Checking the amount of electrical current drawn from the power socket
  • Manual test of the correct device function (AUTOTEST)

To increase work productivity and efficiency REVEXmax together with its software ILLKO Studio is equipped with appliance database feature:

  • Appliance identification using an ID code (barcode)
  • Possibility to transfer the appliance database or selected database parts from PC to REVEXmax
  • Possibility to update the appliance database from the REVEXmax to the PC
  • Possibility to use automatic workflows during testing
  • Manual evaluation of the inspection results directly in the device
  • Possibility to individually modify inspection result forms in PC

Package content

  • REVEXmax W measuring device
  • ILLKO Studio PC software license (the license is loaded in the device)
  • P2011 - safety connecting cable (banana/banana), black, length 2 m
  • P3011 - safety measuring tip, black
  • P2014 - safety connecting cable (banana/banana), red, length 2 m
  • P4016 - crocodile clip, red
  • USB cable
  • Spare fuse F 16 A / 500 V, 32 x 6.3 mm, 2 pcs
  • User´s manual
  • Calibration certificate
  • Warranty card

 

Generally

  • Power supply: 230 V ± 10% / 50 Hz
  • Display: graphic OLED
  • Protection class: II (double insulation)
  • Overvoltage category: II / 300 V
  • Pollution degree: 2
  • Protection: IP 20
  • PC Connection: USB
  • Memory: 16,000 appliance
  • Dimensions: 250 x 195 x 50 mm
  • Weight: about 1.75 kg
  • Working temperature: 0 ÷ 40 °C

Protective conductor resistance

  • Range: 0.00 Ω ÷ 10.00 Ω
  • Open circuit measuring voltage: 5 ÷ 12 V AC
  • Measuring current (R ≤ 1 Ω): ≥ 1 A AC
  • Resistance compensation of test leads: yes (up to 2 Ω)

Insulation resistance

  • Range: 0.20 MΩ ÷ 100.0 MΩ
  • Nominal voltage: 250 V / 500 V DC
  • Nominal current: ≥ 1 mA DC

Leakage current – direct method

  • Range: 0.00 mA ÷ 10.00 mA

Leakage current – differential method

  • Range: 0.00 mA ÷ 10.00 mA

Leakage current – substitute method

  • Range: 0.00 mA ÷ 20.00 mA
  • Open circuit measuring voltage: 230 V ± 15% / 50 Hz
  • Max. measuring current: 3.5 mA

Leakage current through external measuring device

  • Range: 0.00 mA ÷ 20.00 mA

Apparent power consumption

  • Range: 10 VA ÷ 3680 VA

Network Voltage

  • Range: 207.0 V ÷ 253.0 V

Current passing through the appliance

  • Range: 0.0 A ÷ 16.0 A

Current measured by a clamp meter

  • Range: 0.0 A ÷ 100.0 A

Apparent power measured by a clamp meter

  • Range: 10 VA ÷ 9999 VA

Voltage

  • Range: 0.0 V ÷ 130.0 V DC
  • Range: 0.0 V ÷ 90.0 V AC
  • Input resistance: 1.1 MΩ

Welding voltage

  • Range (RMS): 0.0 V ÷ 130.0 V DC / 0.0 V ÷ 90.0 V AC
  • Range (peak): 30.0 V ÷ 130.0 V DC / 30.0 V ÷ 90.0 V AC
  • Input resistance: 200 Ω ÷ 5 kΩ

 

User´s Manual